CMU-K20/K21/K22/K23/K24
Signalling: Connection Control
1115.6088.12 4.159
E-12
Multi Slot
The
Multi Slot
section determines the main timeslot and the way the levels are
defined.
Level Mode
Individual
levels in all downlink channels or distinction between
Used/Unused
timeslots. If
Used/Unused
is selected the
Used
Timeslot Level
and the
Unused Timeslot Level
can be defined in
the
Traffic Channel
section.
Individual
levels are defined relative to
the
Reference Level
defined below.
Reference Level
Reference value for the individual downlink (BS) signal levels. The
reference level is available only if the
Level Mode
is set to
Individual
(see above).
Main Timeslot
Timeslot used for signalling. The main timeslot can not be
switched off in both the downlink and uplink; see
Slot
Configuration
below.
Remote control
CONFigure:BSSignal[:CSWitched][:TCH]:MSLot:LMODe
CONFigure:BSSignal[:CSWitched][:TCH]:MSLot:RLEVel
PROCedure:BSSignal[:CSWitched][:TCH]:MSLot:RLEVel
CONFigure:BSSignal[:CSWitched][:TCH]:MSLot:MTIMeslot
PROCedure:SIGNalling[:CSWitched][:TCH]:MSLot:SCONfig
Slot
Configuration
Table of all used and unused timeslots (GSM timeslots 0 to 7) in the downlink and
the uplink. The
Main Timeslot
is always active (enabled) in both the downlink and
uplink direction. The boxes enable (if checked) or disable the other timeslots.
Important Note:
The CMU can transmit signals in enabled as well as in
disabled downlink timeslots. Enabling a downlink timeslot
means that the MS is instructed to listen to a signal in this
timeslot.
Downlink Level (BS)
RF levels in all downlink timeslots (RF signal transmitted by
the BS/CMU). If the
Level Mode
is set to
Used/Unused,
the
Used Timeslot Level
and the
Unused Timeslot Level
from the
Traffic Channel
section are entered and the downlink levels in
the
Slot Configuration
table can not be edited. If the
Level
Mode
is set to
Individual,
all downlink levels are defined
relative to the
Reference Level
.
If the
BCCH and TCH
mode is active (see
Mode
softkey on p.
4.163
), slot 0 of the downlink signal is reserved for the BCCH
and slots 1 and 7 are not available for the traffic channel.
BCCH
is indicated instead of the level and the enable
switches for slots 0, 1 and 7 are grayed.
Uplink PCL (MS)
Transmitter output power of the MS in all active (enabled)
uplink timeslots. The MS transmitter output power is selected
in PCL (Power Control Level) units; the corresponding
absolute power value is also indicated; see section
Limit
Values for Average Burst Power (Power Configuration –
Limits)
on page
4.112
. If a timeslot is disabled, the
corresponding output power is automatically switched
Off
.
Note:
The number of downlink and uplink channels must be compatible with
the multislot class of the MS under test; see
Table 4-10
on p.
4.149
.
Remote control
CONFigure:BSSignal[:CSWitched][:TCH]:MSLot:SCONfig:UUNused
CONFigure:BSSignal[:CSWitched][:TCH]:MSLot:SCONfig:INDividual
CONFigure:MSSignal[:CSWitched][:TCH]:MSLot:SCONfig
PROCedure:SIGNalling[:CSWitched][:TCH]:MSLot:SCONfig
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...