Messages in the dialog
An information message (or error message) is displayed if one of the following hap-
pens:
●
One of the selected calibration kits is described by ideal kit parameters or typical
values.
●
One of the selected calibration kits does not contain all standards that are required
for the chosen calibration type.
●
Different connector types are defined at the ports but the selected calibration type
requires uniform connectors.
●
A cal kit standard, cal unit or power meter does not cover the entire calibrated fre-
quency range.
●
A combination of cal units and cal kits is configured in a channel
Calibration Type tab
The "Calibration Type" tab allows you to configure the calibration setup (i.e. a set of
calibration definitions) for the channel selected in the title area of the calibration setup
dialog.
Use Power Cal
If checked, you can select the power calibrations to be performed together with the
system error correction.
The default setting and the selected power cals depend on the active channel mode.
E.g. for a standard (S-parameter) channel a
is available.
Source Leveling/Src. Lev. Method
Adds a
of the related port to the current calibration definition,
which comes in handy for SMARTer calibrations (see
SMARTerCal with scalar power calibration"
When you run the
performed.
"Src. Lev. Method" is equivalent to "Pwr Cal Method" on the "Power Cal Settings" tab
(see
"Flatness Cal – Pwr Cal Method"
Remote command:
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect:METHod
Connector / Gender
Defines the connector types and genders of the ports to be calibrated. For symmetric
(sexless) connector types (e.g. 7 mm / PC7), "Gender" is unavailable.
If "Default Conn." is active, the "Default Connector" is set in the "Default for all Chan-
nels" section (to the left of the
table). If "Same Gen." is active, the gen-
ders at all ports are always adjusted to the current selection.
User-defined connectors can be added or removed in the
Cal softtool