GUI reference
R&S
®
ZNA
557
User Manual 1178.6462.02 ─ 20
Use "Detect Port Assignment" to detect the current port assignment (after reconnecting
the calibration unit).
Measurement
Selects the measurement to be performed in advanced mode and runs it
Characterization
Selects the calibration unit characterization to be used for the validation.
By default the factory characterization is used.
Remote command:
CALCulate:CALValidate:CHARacteriza
Calibration Standard
Selects the calibration standard to be measured in advanced mode.
Validation Limits
The "Limits..." button opens the "Cal Validation Limits" dialog that allows you to specify
the maximum tolerable deviations from the corresponding S-parameters of the
selected reference
Defaults are ±0.2
dB for magnitude and ±2° for phase deviations.
Remote command:
Cal softtool