Command reference
R&S
®
ZNA
958
User Manual 1178.6462.02 ─ 20
CALCulate:FMODel:SFD<Ph_pt>:PRESet
Restores the default settings of the SFD tool.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Usage:
Event
Manual operation:
See
CALCulate:FMODel:SFD<Ph_pt>:DUT[:STATe]
<Boolean>
For a fixture modeling with the SFD tool, this command allows to specify the ports to
which the test fixture is connected.
This has to be defined before measuring DUT + Fixture (using
).
Suffix:
<Ph_pt>
.
Physical port number
Parameters:
<Boolean>
1 (ON, true) if the test fixture is connected to port <Ph_pt>, 0
(OFF, false) otherwise
Manual operation:
See
CALCulate:FMODel:SFD<Ph_pt>:RUN:RUN
Runs the SFD tool.
Before executing this command, make sure that:
●
The test coupon measurement (using
) finished successfully, or the test coupon data were success-
CALCulate:FMODel:SFD<Ph_pt>:COUPon:
●
The measurement of DUT + test fixture (using
The resulting Touchstone files are written to C:\Users\Public\Documents\Rohde-
Schwarz\ZNA\Embedding. If result files with the same name already exist, they are
overwritten.
Suffix:
<Ph_pt>
.
This suffix is ignored
Usage:
Event
Manual operation:
See
SCPI command reference