MMEMory:LOAD:CKIT:SDATa:WLABel
<ConnectorType>, <CalKitName>,
<KitLabel>, <StandardType>, <StandardLabel>, <TouchstoneFile>[,
<FirstPort>[, <SecondPort>]]
Loads characterization data from the given Touchstone file; similar to existing com-
mand
but supports cal kit addressing
by label
.
Setting parameters:
<ConnectorType>
String parameter containing the name of the connector type.
<CalKitName>
String parameter containing the name of a calibration kit availa-
ble on the analyzer.
<KitLabel>
String parameter containing the label of a calibration kit available
on the analyzer. An empty string means that no label is defined.
<StandardType>
MMTHrough | MFTHrough | FFTHrough | MMLine | MMLine1 |
MMLine2 | MMLine3 | MFLine | MFLine1 | MFLine2 | MFLine3 |
FFLine | FFLine1 | FFLine2 | FFLine3 | MMATten | MFATten |
FFATten | MMSNetwork | MFSNetwork | FFSNetwork | MOPen |
FOPen | MSHort | FSHort | MOSHort | MOSHort1 | MOSHort2 |
MOSHort3 | FOSHort | FOSHort1 | FOSHort2 | FOSHort3 |
MREFlect | FREFlect | MMTCh | FMTCh | MSMatch | FSMatch
Standard types; for a description refer to table
.
<StandardLabel>
String parameter addressing a particular calibration standard by
its label. An empty string means that no label is defined.
<TouchstoneFile>
String parameter to specify the name and directory of the Touch-
stone file to be loaded. A
*.s1p
file must be used for one-port
standards, a
*.s2p
file for two-port standards.
If no path is specified the analyzer searches the current direc-
tory, to be queried with
<FirstPort>
First port number (sufficient for one-port standards). If the port
numbers are omitted, the cal kit data is valid for all ports.
<SecondPort>
Second port number, for two-port standards. If the port numbers
are omitted, the cal kit data is valid for all ports.
Example:
MMEM:LOAD:CKIT:SDAT:WLAB 'N 50 Ohm','Default
Kit','0815',MOPEN,'Test data','test.s1p',1
Load the file Test.s1p from the current directory in order to
define the properties of an Open (m) standard in the cal kit
"Default Kit" with label "0815" for the N 50 Ω connector type.
Assign the label "Test data" and specify that the standard data is
valid for port no. 1 only.
Usage:
Setting only