GUI reference
R&S
®
ZNA
662
User Manual 1178.6462.02 ─ 20
5.14.2.1
Controls on the Offset tab
Overview
This button is available on all "Offset Embed" softtool tabs. It opens the
in the
Port
Physical test port of the analyzer. You can define independent offset parameters for all
ports.
Remote command:
The
<PhyPt>
numeric suffix in the
[SENSe<Ch>:]CORRection:...
commands
identifies the physical port.
Delay / Electrical Length / Mech. Length
Defines the length offset at the selected port as a delay, an electrical length, or a
mechanical length. The three quantities are related by:
and overwrite each other. See also
Chapter 4.6.1.1, "Definition of offset parameters"
Note:
The entered parameters must correspond the actual (one-way) length of the
transmission line. To account for the propagation in both directions, the phase shift of a
reflection parameter due to a given length offset is twice the phase shift of a transmis-
sion parameter. For a numeric example, see
Chapter 4.6.1.6, "Application and effect of
Offset Embed softtool