Command reference
R&S
®
ZNA
1320
User Manual 1178.6462.02 ─ 20
Parameters:
<Boolean>
Enables (ON) or disables (OFF) the receiver power calibration
for the reference waves a
n
.
*RST:
OFF
Example:
See
[SENSe<Ch>:]CORRection:POWer<PhyPt>:ACQuire
[SENSe<Ch>:]CORRection:POWer<PhyPt>:HARMonic:ACQuire
Starts the receiver calibration
●
at physical port <PhyPt>
●
in channel <Ch>
●
at harmonic orders
[SENSe<Ch>:]CORRection:ADVanced:HARMonic:ORDer?
Then stores and applies the calibration data.
Note:
The receiver calibration relies on the source power calibration acquired using
SOURce<Ch>:POWer<PhyPt>:CORRection:HARMonic[:ACQuire]
.
Suffix:
<Ch>
.
Channel number
<PhyPt>
Port number.
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:
Usage:
Event
Manual operation:
See
"Harmonic Orders to Calibrate"
[SENSe<Ch>:]CORRection:POWer<PhyPt>:IMODulation:ACQuire
[<CalOnlyPortFreq>]
Starts the receiver calibration (3
rd
power calibration step for intermodulation measure-
ments), stores and applies the calibration data.
Note:
The receiver calibration relies on the source power calibration acquired in step
nos. 1 and 2.
[SENSe<Ch>:]CORRection:POWer<PhyPt>:IMODulation:ACQuire
is enabled
only after steps 1 and 2 have been carried out (
CORRection:IMODulation:LTONe[:ACQuire]
CORRection:IMODulation:UTONe[:ACQuire]
).
Suffix:
<Ch>
.
Calibrated channel number
SCPI command reference