Annexes
R&S
®
ZNA
1652
User Manual 1178.6462.02 ─ 20
Table 12-5: Voltage/Current Measurement (Variant 03 only)
Range
Resolution
Voltage
RFFE 1 and 2
(VIO/DATA/CLK)
0 V to +3 V
100 μV
GPIO 1 to 10
-5 V to +10 V
100 μV
Current*
RFFE 1 and 2 (VIO/DATA/CLK), GPIO 1 to 8
10 Ω source resistance (shunt)
–20 mA to +20 mA
10 μA
100 Ω source resistance (shunt)
–2 mA to +2 mA
1 μA
1 kΩ source resistance (shunt)
–200 μA to +200 μA
100 nA
10 kΩ source resistance (shunt)
–20 μA to +20 μA
10 nA
100 kΩ source resistance (shunt)
–2 μA to +2 μA
1 nA
GPIO 9 and 10
–100 mA to +100 mA
10 μA
*
the current values are valid if the GPIO voltages are within -5
V to +9
V
12.4
Showroom mode
In "Showroom Mode" the R&S
ZNA uses a configurable recall set whenever the device
is (re-)started or the [Preset] key is pressed. However, it can only be enabled by
directly editing the Windows registry.
1. Run the R&S
ZNA and configure it as required for the intended showroom opera-
tion.
2. Save the configuration as described in
Chapter 5.15.1, "Recall sets tab"
on page 707. If necessary, move the recall set file (
*.znxml|*.znx
) to the appro-
priate location.
3. Under the registry key
HKEY_LOCAL_MACHINE/SOFTWARE/Rohde-Schwarz/ZNA
, create a new subkey
ShowroomMode
.
4. Within the
ShowroomMode
-subkey, create the string value
ShowroomSetup
.
Showroom mode