Command reference
R&S
®
ZNA
1334
User Manual 1178.6462.02 ─ 20
LO2 – local oscillator signal no. 2, for 2-stage mixer measure-
ments (
[SENSe<Ch>:]FREQuency:CONVersion:MIXer:
IF – mixer output signal
Range:
Depending on the instrument model [Hz]. The incre-
ment (parameters UP or DOWN) is 0.1 kHz.
*RST:
Minimum of the analyzer's frequency range
<FixedFrequency>
Default unit: Hz
Example:
The following example assume a four-port analyzer with inde-
pendent source ports no. 1 and 3. Moreover, an external genera-
tor Gen 1 must be connected to the analyzer and configured.
*RST; FREQ:CONV:MIX:STAG 2
Select a mixer measurement with two mixer stages.
FREQ:CONV:MIX:RFSource 1; IFSource 2
Select the analyzer port 1 as a source port for the RF signal,
port 2 as a receive port for the IF signal.
FREQ:CONV:MIX:LOSource1 PORT, 3; LOSource2 GEN,
1
Select port 3 as a source port for the local oscillator LO1, port 3
as a source for LO 2.
FREQ:CONV:MIX:RFMultiplier 2, 1; LOMultiplier1
2, 1 ; LOMultiplier2 2, 1
Define frequency conversion factors 2 for the RF signals and the
LO signals.
FREQ:STARt 1GHz; STOP 2 GHZ
Select a sweep range between 1 and 2 GHz.
FREQ:CONV:MIX:FIXed1 LO1; FIXed2 LO2
Assign fixed frequency settings to the two LO signals.
FREQ:CONV:MIX:FUNDamental RF; MFFixed LO1, 10
MHz; MFFixed LO2, 10 MHz
Assign the channel base frequency (sweep range) to the RF sig-
nal, fixed frequencies of 10 MHz to both LO signals.
FREQ:CONV:MIX:TFRequency1 DCUP; TFRequency2
UCON
Select down-conversion (USB) at the first mixer-stage, up-con-
version (USB) at the second mixer stage.
SOUR:FREQ:CONV:MIX:PMODe RF, FUNDamental; PMODE
LO1, FIXed; PMODe LO2, FIXed; PMODe IF,
FUNDamental
Assign the channel base power to the RF and IF signals, a fixed
power to the LO signals.
SOUR:FREQ:CONV:MIX:PMFixed LO1, -20; PMFixed
LO2, -10
Define the fixed powers of the LO signals.
Manual operation:
See
SCPI command reference