GUI reference
R&S
®
ZNA
551
User Manual 1178.6462.02 ─ 20
left = "Ref. Receiver"
right = "Power Meter"
Start Cal Sweep
Start the calibration sweeps for the selected port and power calibration settings and
close the dialog. The calibration is performed as described in
Use the
"Pwr Cal Settings" to modify the calibration procedure.
Remote command:
SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect[:ACQuire]
5.11.1.6
Calibrate All dialog
The "Calibrate All" dialog guides you through the setup of complex calibration scenar-
ios, where the set of calibrations to be performed for each port group and channel is
freely configurable. The calibration configuration is persisted and can be recalled.
Background and related information
See
Chapter 4.5.8, "Parallel calibration of multiple channels"
Access:
Channel – [Cal] > "Start Cal" > "Calibrate All"
–
"Con Start..."
Cal softtool