Command reference
R&S
®
ZNA
1099
User Manual 1178.6462.02 ─ 20
FPORts (or omitted)
Standard port sequence (network port 1 towards VNA, network
port 2 towards DUT)
IPORts | SGATes
Inverted port sequence (network port 2 towards VNA, network
port 1 towards DUT)
<arg1>
<block_data>
Content of a Touchstone file (
*.s2p
) in IEEE488.2
.
Usage:
Setting only
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<PhyPt>:
PARameters:C<Cmp>
<CircuitModel>, <Capacitance>
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<PhyPt>:
PARameters:C<Cmp>?
<CircuitModel>
Specifies the capacitance value C<Cmp> in the different circuit models for single
ended port embedding.
In the query form, the <Capacitance> parameter must be omitted. The command
returns the capacitance value for the specified circuit model.
Suffix:
<Ch>
.
Channel number
<PhyPt>
Physical port number
<Cmp>
Number of capacitance in circuit model. The total number of
capacitances depends on the selected circuit model.
Parameters:
<Capacitance>
Capacitance C<Cmp> for the specified circuit model.
Range:
-1mF to 1 mF.
Increment: 1 fF (1E-15 F)
*RST:
1 pF (1E-12 F)
Default unit: F
Parameters for setting and query:
<CircuitModel>
CSL | LSC | CSC | SLC | SCL | SCC | SHLC | GSL | LSG |
GSG | SLG | SGL | SGG
Possible circuit models (character data); see
Example:
*RST; :CALC:TRAN:VNET:SEND:DEEM:PAR:C2? CSC
Query the default capacitance C2 for the Serial C, shunt C cir-
cuit model. The response is 1E-012 (1 pF).
CALC:TRAN:VNET:SEND:DEEM:PAR:C2 CSC, 2.2E-12
Increase the capacitance to 2.2 pF.
Manual operation:
See
SCPI command reference