Command reference
R&S
®
ZNA
1402
User Manual 1178.6462.02 ─ 20
Manual operation:
See
"All Partial Meas'ments / First Partial Meas'ment"
[SENSe<Ch>:]SWEep:FPOWer
<Mode>
Selects the power sweep mode.
Suffix:
<Ch>
.
Parameters:
<Mode>
OFF | ON | FAST
OFF
Regular power sweep
ON
Fast power sweep
FAST
Pulsed power sweep
Manual operation:
See
[SENSe<Ch>:]SWEep:GENeration
<mode>
Sets/gets the sweep mode for linear frequency sweeps.
Suffix:
<Ch>
.
Channel number
Parameters:
<mode>
STEPped
The R&S
ZNA only supports the stepped mode, i.e, the fre-
quency is changed stepwise and sampling is performed at a
tuned frequency for each measurement point.
*RST:
STEPped
[SENSe<Ch>:]SWEep:POINts
<SweepPoint>
Defines the total number of measurement points per sweep ("Number of Points").
Values between 1 and 100,001 can be set.
Suffix:
<Ch>
.
Channel number 1
Parameters:
<SweepPoint>
Number of points per sweep
Range:
see above
*RST:
201
SCPI command reference