GUI reference
R&S
®
ZNA
498
User Manual 1178.6462.02 ─ 20
Tip:
Overlapping limit line and ripple limit line segments are not displayed when a
point-based x-axis is active; see
Chapter 4.4.1.1, "Rules for limit line definition"
Remote command:
[SENSe<Ch>:]FREQuency:SEGMent:AXIS
Reverse
Reverses the direction of the sweep.
Remote command:
Fast Power Sweep
This control is only enabled, if sweep type
is selected.
The alternative power swep methods are most suitable for compression measurements
on freqency-converting DUTs with embedded (and drifting) LO. For this kind of mea-
surements, "Track LO" alone does not help, because not only the frequency, but also
the phase can vary between sweep points, which makes power level measurements
unreliable.
Off
Regular power sweep (stepped mode). For each sweep point, mea-
surement takes place after the power level has settled.
Sweep Softtool