Command reference
R&S
®
ZNA
1132
User Manual 1178.6462.02 ─ 20
Manual operation:
See
CONTrol<Ch>:SEQuence<Nr>:DELay
<Delay>
For unsegmented sweeps this command allows to introduce delays between subse-
quent sequence steps and between the final sequence step and the sweep (segment)
start.
The command/switch sequences are defined using
CONTrol<Ch>:SEQuence<Nr>:GPIO<Port>:
.
Suffix:
<Ch>
.
Channel number
<Nr>
Sequence number, defining the order in which the commands
shall be executed.
For every channel and RFFE interface, the sequence numbers
must be consecutive, starting at 1; reusing a sequence number
overwrites a previously defined command.
Parameters:
<Delay>
Delay time
Default unit: s
Manual operation:
See
"Wait (Sweep Sequencer Table)"
CONTrol<Ch>:SEQuence<Nr>:GPIO<Port>:VOLTage
<Voltage>
For unsegmented sweeps this command allows to define the GPIO voltage settings to
be applied at the start of each sweep.
The GPIO ports can be enabled/disabled using
. Complementary RFFE commands can be defined using
SEQuence<Nr>:RFFE<Bus>:COMMand:DATA
.
Suffix:
<Ch>
.
Channel number
<Nr>
Sequence number, defining the order in which the switches/
commands shall be executed.
For every channel and GPIO port, the sequence numbers must
be consecutive, starting at 1; reusing a sequence number over-
writes a previous voltage setting.
Delays between subsequent settings/command and - finally - the
start of the respective sweep segment can be introduced using
CONTrol<Ch>:SEQuence<Nr>:DELay
<Port>
GPIO port number
SCPI command reference