GUI reference
R&S
®
ZNA
465
User Manual 1178.6462.02 ─ 20
Target Format
Description
Formula
"Phase unwrap"
Unwrapped phase of z comprising
the complete number of 360°
phase rotations [°]
Ф(z) = φ (z) + 2k·360°
"Real"
Real part of z
Re(z) = x
"Imag"
Imaginary part of z
Im(z) = y
"SWR"
(Voltage) Standing Wave Ratio
SWR = (1 + |z|) / (1 – |z|)
"Default"
Identical to trace format.
Note
: the Smith and Polar traces
use "Lin Mag" as the default for-
mat for target value.
-
Remote command:
CALCulate<Chn>:MARKer<Mk>:SEARch:FORMat
Target Search
Activates the search and sets the active marker to the defined target value. If the target
value occurs at several stimulus values, the marker is placed to the search result with
the smallest stimulus value. The other measurement points with the same target value
can be located using the "Search Right" function.
If the target is not found (e.g. because the active trace does not contain the target
value), then the active marker is not moved away from its original position.
Remote command:
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute
TARGet
CALCulate<Chn>:MARKer<Mk>:FUNCtion:RESult?
Search Left/Search Right
Searches the
to the left/right of the active marker's stimulus value within
Chapter 5.7.3.2, "Search Range dialog"
If the search is successful, the active marker is moved to the next smaller/larger stimu-
lus value with this target value. Use "Search Left"/"Search Right" repeatedly to locate
the other ones.
Remote command:
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute
LTARget
CALCulate<Chn>:MARKer<Mk>:FUNCtion:RESult?
Search Mode
Displays and sets the current marker search mode.
Select one of the predefined max, min, peak, or target searches or select "Sweep Pro-
gress" to track the position of the sweep cursor.
Remote command:
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute
Marker softtool