Self test error
In this case enable error logging for external devices (see
for self test error codes of the
device.
●
–
The device driver does not match exactly the device type. Probably the device
cannot be controlled properly. Measurements results are questionable.
Remote command:
SYSTem:COMMunicate:RDEVice:GENerator<Gen>:DEFine
SYSTem:COMMunicate:RDEVice:GENerator:DELete
SYSTem:COMMunicate:RDEVice:GENerator<Gen>:CATalog?
SYSTem:COMMunicate:RDEVice:GENerator<Gen>:DEFine
SYSTem:COMMunicate:RDEVice:GENerator<Gen>:SEPMode
SYSTem:COMMunicate:RDEVice:GENerator<Gen>:SEPower
Scan Instruments
Performs an autodetection of "Known Devices" on the supported interfaces.
Note:
Unintentional switchover to remote control
When using the NI-VISA library, ensure that the network analyzer itself is not listed as
a network device in the Measurement & Automation Explorer. Otherwise, "Scan Instru-
ments" will send an identification query (*IDN?), causing the analyzer to close the
"External Generators" dialog (without executing "Scan Instruments") and to activate the
remote screen.
Remote command:
n/a
LAN Detection
Activate "LAN Detection" to include the LAN interface in the autodetection sequence
(see
).
Note:
"LAN detection" will only work for external devices sharing the same IP subnet
with the R&S
ZNA.
As a prerequisite, the R&S
ZNA must accept incoming connections on UDP port 2473.
If necessary, add a corresponding inbound rule to your firewall settings.
Remote command:
n/a
Add Device
Adds a new instrument to the list of "Configured Devices". In the "Add External Gener-
ator" dialog opened, you can specify the properties of the new instrument and the con-
nection.
Setup softtool