Command reference
R&S
®
ZNA
963
User Manual 1178.6462.02 ─ 20
Parameters:
<Boolean>
ON | OFF | 1 | 0
*RST:
0
Manual operation:
See
CALCulate<Chn>:IAVerage:MODE
<Mode>
Selects the quantities to be averaged if infinite averaging is enabled
(
CALCulate<Chn>:IAVerage[:STATe] ON
.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Mode>
MAGPhase | REIMag
Selects the averaging mode.
MAGPhase
Averaging of magnitude and phase (default) of the complex
trace value
REIMag
Averaging of real and imaginary part of the complex trace value
Manual operation:
See
7.3.1.12
CALCulate:LDEViation...
CALCulate<Chn>:LDEViation:AUTO
<Activate>
Initiates a (re-)calculation of the linearity deviation correction factors and applies them
to the active trace.
Suffix:
<Chn>
.
Channel number used to identify the active trace. If unspecified
the numeric suffix is set to 1.
Setting parameters:
<Activate>
ONCE
Initiates the calculation
Example:
*RST; CALC:FORM PHAS
Reset the analyzer; assign the phase of the transmission param-
eter S21 to the default trace.
CALC:LDEV:AUTO ONCE
Calculate correction factors and apply them to the default trace.
CALC:LDEV:SLOP?; ELEN?; CONS?
Query the values of the correction factors.
CALC:LDEV:MODE OFF
Disable the correction.
Usage:
Setting only
SCPI command reference