GUI reference
R&S
®
ZNA
761
User Manual 1178.6462.02 ─ 20
5.17.2.13
X Axis tab
The "X Axis" tab allows you to define the x axis scaling of the active diagram. The scal-
ing logic is the same as for all other diagram types, see
Access:
System – [Applic] > "TDR" > "X Axis"
●
This tab is only visible if the
ZNA-K20
is available.
●
If the active trace is represented as an eye diagram, the "X Axis Settings" are
grayed out, as shown above.
If the active trace is represented as a rise time or skew diagram, the "X Axis Set-
tings" allow you to adjust the center/span or start/stop, no matter if the X axis rep-
resents "Time" or "Distance" (see
5.17.3
Distance to Fault application
With option K2, the R&S
ZNA can locate faults and discontinuities on cables and trans-
mission lines. The measured S
ii
-parameter trace is mathematically converted to the
time domain, and represented in the distance domain. The x-axis of the DtF trace
(derived from S
ii
) shows the distance from the calibrated reference plane; faults appear
as peaks above a certain level.
If option R&S
ZNA-K2 is not installed on the R&S
ZNA, the "Distance to Fault" applica-
tion is not available.
To achieve the best distance resolution, set the sweep span as large as possible and
use the
to adjust the number of sweep points.
Applic softtool