Command reference
R&S
®
ZNA
1307
User Manual 1178.6462.02 ─ 20
[SENSe<Ch>:]CORRection:DATA:PARameter<Sfk>?
[<SfkSettingType>[, <Index>]]
[SENSe<Ch>:]CORRection:DATA:PARameter<Sfk>:PORT<PhyPt>?
[<SfkSettingType>[, <Index>]]
Gets the settings of active system error correction no. <Sfk> (or of the factory calibra-
tion, if no channel calibration is active).
Suffix:
<Ch>
.
Number of the calibrated channel.
<Sfk>
Number of the system error correction.
Less or equal than the total number of active system error cor-
rections for the related channel (see
]CORRection:DATA:PARameter:COUNt?
<PhyPt>
Physical port number, used to select a specific frequency axis in
arbitrary mode (with option R&S
ZNA-K4).
Query parameters:
<SfkSettingType>
ACAL | STARt | STOP | POINts | SPOWer | STYPe |
BANDwidth | PDLY | RATTenuation | TYPE | PORTs | SPORt |
THRoughs | TSTamp | LTSTamp | TVNA | MVNA | MTESt |
CKIT | FSMode
The requested setting.
If no <SfkSettingType> is specified, the values for STARt, STOP,
POINts, SPOWer and STYPe are returned
ACAL
1 for automatic calibrations, 0 for manual calibrations
STARt
Start frequency (or CW frequency, if no frequency sweep is
active)
STOP
Stop frequency (or CW frequency, if no frequency sweep is
active)
POINts
Number of points
SPOWer
Source power (or stop power, for power sweeps)
STYPe
Sweep type or grid (LIN, LOG, SEGM)
BANDwidth
Measurement bandwidth
PDLY
Point delay (Meas Delay) for OSM
RATTenuation
Receiver attenuations: comma-separated list containing a value
pair <port no.>,<rec. att.> (float,integer) for each of the involved
PORTs
SCPI command reference