GUI reference
R&S
®
ZNA
788
User Manual 1178.6462.02 ─ 20
Global Settings
The two buttons reset all directory settings (e.g. the directories for storing trace data,
limit lines, calibration data...) and all settings in the "Printer Setup" dialog to default val-
ues. See
Chapter 5.15.4, "Printer Setup dialog"
Remote command:
n/a
Preset Configuration
Specifies whether the preset action performs a factory preset or restores a user preset
file. A user preset file is an arbitrary recall set file (
*.znxml|*.znx
), that can be cre-
ated from the active setup using System – [File
Print] > "Save...". If the current user
preset file is not found (e.g. because it was deleted or moved), System – [Preset] ini-
tiates a factory preset.
The preset configuration applies to:
●
the System – [Preset] hardkey (if
on page 843 is selected)
●
the
●
the remote commands
*RST
and
SYSTem:PRESet
(if "Align *RST to User Defined
Furthermore, the user preset file is used as a template, whenever
are created from the GUI.
Remote command:
SYSTem:PRESet:USER:NAME
SYSTem:PRESet:USER[:STATe]
Start in Preset
If this option is active, the R&S
ZNA FW always starts with the configured
. Otherwise it starts with the previous state, i.e. the settings that were persis-
ted during the previous shutdown.
Remote command:
Calibration tab
Provides general system error correction (calibration) settings.
Setup softtool