Command reference
R&S
®
ZNA
1515
User Manual 1178.6462.02 ─ 20
SYSTem:PRESet:USER:NAME
<RecallSetFile>
Specifies the name of a recall set file (
*.znxml|*.znx
) to be used for a user-defined
preset.
Parameters:
<RecallSetFile>
String parameter to specify the absolute or relative path of the
recall set file to be loaded. Relative paths are evaluated relative
to the current directory (see
Example:
See
Manual operation:
See
SYSTem:SETTings:UPDate
<Activate>
Initiates an immediate update of the channel or trace settings.
The command has an effect if the analyzer operates in single sweep mode
(
OFF
) and if the display update is switched off
OFF
). In this scenario, a change of the channel or trace
settings is usually not taken into account immediately. The analyzer waits until the end
of the current sweep sequence and changes all settings made during the last sweep
period when the next single sweep sequence is initiated. Several settings can be made
en bloc, which generally saves time.
SYSTem:SETTings:UPDate ONCE
causes the analyzer to apply the settings at once
without waiting for the end of the current single sweep sequence. The command has
no effect in continuous sweep mode or if the display update is switched on.
The settings are also updated when the continuous sweep mode is activated
(
ON
).
Setting parameters:
<Activate>
ONCE
Causes an immediate update of the settings.
Example:
INIT:CONT OFF
Activate single sweep mode.
SYST:SETT:UPD ONCE
Update the settings made during the current single sweep
period.
Usage:
Setting only
SYSTem:SHUTdown
[<Unit>]
Performs a shutdown or restart of the FW or OS.
If the optional parameter
<Unit>
is omitted, Windows is shutdown afer a time-out
period of 10 seconds.
Setting parameters:
<Unit>
HALT | REBoot | ABORt | CLOSe | RESTart
SCPI command reference