Command reference
R&S
®
ZNA
1530
User Manual 1178.6462.02 ─ 20
TRIGger:CHANnel<Ch>:AUXiliary<n>:INTerval
<Type>
Selects the sweep interval to be indicated at the re:ALClated Trigger Out connector.
Suffix:
<Ch>
.
Channel number
<n>
1 = Trigger Out A, ..., 4 = Trigger Out D
Parameters:
<Type>
NONE | SWEep | SEGMent | PPOint | POINt | ACQuisition |
PGENerator
NONE
No trigger out signal for this channel and Trigger Out connector.
SWEep | SEGMent | PPOint | POINt
A trigger out signal is generated for every sweep, segment, par-
tial measurement or sweep point, respectively.
ACQuisition
A trigger out signal is generated for every data acquisition phase
PGENerator
For pulse generator pulses, pulses with identical or opposite
polarity can be generated as trigger out signal
*RST:
SWEep
Example:
TRIGger:CHANnel1:AUXiliary1:STYPe POSitive
Generate pulses with positive polarity at Trigger Out A.
TRIGger:CHANnel:AUXiliary:INTerval SWEep
Send one output trigger pulse per sweep.
TRIGger:CHANnel:AUXiliary:POSition BEFore
TRIGger:CHANnel1:AUXiliary1:DELay 1 us
TRIGger:CHANnel:AUXiliary:DURation 10 us
Send the output trigger 1
μs before sweep start, and with 10
μs
pulse width.
Options:
R&S
ZNA-B91
Manual operation:
See
TRIGger:CHANnel<Ch>:AUXiliary<n>:POSition
<Type>
If pulses are used as trigger out signals (see
), the pulses can either indicate the "Start" or "End" of the rela-
ted sweep interval (see
TRIGger:CHANnel<Ch>:AUXiliary<n>:INTerval
Suffix:
<Ch>
.
Channel number
<n>
1 = Trigger Out A, ..., 4 = Trigger Out D
Parameters:
<Type>
BEFore | AFTer