GUI reference
R&S
®
ZNA
409
User Manual 1178.6462.02 ─ 20
Remote command:
CALCulate<Chn>:FILTer[:GATE]:TIME:CENTer
CALCulate<Chn>:FILTer[:GATE]:TIME:SPAN
CALCulate<Chn>:FILTer[:GATE]:TIME:STARt
CALCulate<Chn>:FILTer[:GATE]:TIME:STOP
Show Range Lines
Displays or hides two red lines indicating the start and stop of the time gate in a time
domain diagram.
Remote command:
CALCulate<Chn>:FILTer[:GATE]:TIME:SHOW
Bandpass / Notch
The filter type defines what happens to the data in the specific time region.
●
A "Bandpass" filter passes all information in the specified time region and rejects
everything else.
●
A "Notch" filter rejects all information in the specified time region and passes every-
thing else.
Remote command:
CALCulate<Chn>:FILTer[:GATE]:TIME[:TYPE]
Shape
Selects a gate shape which the R&S
ZNA uses to filter the trace in the time domain.
The drop-down list visualizes how the time gate will affect a constant function after
transformation back into the frequency domain. The selected window is applied to the
active trace. The two red vertical lines represent the "Start" and "Stop" values defining
the size of the time gate.
See also
Remote command:
CALCulate<Chn>:FILTer[:GATE]:TIME:SHAPe
CALCulate<Chn>:FILTer[:GATE]:TIME:WINDow
Side Lobe Level
Defines the side lobe suppression for an "Arbitrary Gate Shape (Dolph-Chebychev)"
gate. The entered value is the ratio of the power of the central lobe to the power of the
first side lobe in dB.
Remote command:
CALCulate<Chn>:FILTer[:GATE]:TIME:DCHebyshev
5.5.7
Distance to Fault tab
This tab is only visible, if option R&S
ZNA-K2 is installed on the R&S
ZNA. It replicates
the "Distance to Fault" tab of the "Distance to Fault" application. See
"Distance to Fault application"
Traces softtool