Command reference
R&S
®
ZNA
938
User Manual 1178.6462.02 ─ 20
CALCulate:FMODel:DELT<Ph_pt>:SWEep:CONTrol:STEP?
For the Delta-L measurement, the firmware can either use a default frequency step
size, or the active channel setting (see
CALCulate:FMODel:DELT<Ph_pt>:SWEep:
).
This query returns the resulting frequency step size.
Suffix:
<Ph_pt>
.
This suffix is ignored
Usage:
Query only
Options:
R&S
ZNA-K231
Manual operation:
See
CALCulate:FMODel:DELT<Ph_pt>:SWEep:FREQuency:MAXimum
<Frequency>
CALCulate:FMODel:DELT<Ph_pt>:SWEep:FREQuency:MINimum
<Frequency>
Defines the sweep range of the Delta-L measurement channel. The entered MAXimum
and MINimum values are set as
and
, respectively.
Furthermore, the VNA collects data for the LPI computation a the MAXimum and MINi-
mum frequencies (in addition to
).
Suffix:
<Ph_pt>
.
This suffix is ignored
Parameters:
<Frequency>
Frequency values
The default MINimum frequency is the standard Delta-L start fre-
quency of 10
MHz. The default MAXimum frequency is 20
GHz.
Default unit: Hz
Options:
R&S
ZNA-K231
Manual operation:
See
CALCulate:FMODel:DELT<Ph_pt>:TCONfig
<Boolean>
Defines how the analyzer firmware sets up the diagram area for the Delta-L measure-
ment.
Suffix:
<Ph_pt>
.
This suffix is ignored
Parameters:
<Boolean>
ON (1)
Use the recommended trace configuration, i.e. display each of
the traces in a separate diagram.
SCPI command reference