GUI reference
R&S
®
ZNA
812
User Manual 1178.6462.02 ─ 20
Reference Source
Selects the internal or an external reference clock signal for synchronization.
"Internal"
The analyzer synchronizes to its internal reference oscillator.
The synchronized internal 10 MHz reference clock is also made avail-
able at the rear panel Reference Out BNC connector. In addition, a
synchronized 100 MHz reference signal can be provided at the Refer-
ence In/Out SMA connector (see
on page 813). Use these signals to synchronize other devices.
"External
(BNC)"
The analyzer synchronizes to an external reference clock via its rear
panel Reference In BNC connector.
The external reference signal has to meet the specifications of the
data sheet; its frequency must be specified in the
The synchronized internal 10 MHz reference clock is also made avail-
able at the Reference Out BNC connector. In addition, a synchron-
ized 100 MHz reference signal can be provided at the Reference
In/Out SMA connector (see
on page 813). Use these signals to synchronize other devices.
"External
(SMA)"
The analyzer synchronizes to an external 100
MHz or 1
GHz refer-
ence clock via its rear panel Reference In/Out SMA connector.
No reference signal is provided to the Reference Out BNC connector.
Remote command:
[SENSe<Ch>:]ROSCillator[:SOURce]
Reference Out (BNC)
The outgoing reference signal the analyzer provides at its rear panel Reference Out
BNC connector depends on the selected
Setup softtool