Command reference
R&S
®
ZNA
1249
User Manual 1178.6462.02 ─ 20
Example:
MMEM:LOAD:CKIT 'C:\Users\Public\
Documents\Rohde-Schwarz\ZNA\
Calibration\Kits\New Kit 1.calkit'
Load the previously created cal kit file
New Kit 1.calkit
from the default cal kit directory.
CORR:CKIT:SEL 'N 50 Ohm', 'New Kit 1'
Assign the imported kit to the N 50 Ω connector type, assuming
that the cal kit name stored in
New Kit 1.calkit
reads
New
Kit 1
.
Manual operation:
See
[SENSe:]CORRection:CKIT:STANdard:CATalog?
<CalKitName>
Returns a list of all standards in a given calibration kit.
Query parameters:
<CalKitName>
Example:
See
[SENSe<Ch>:]CORRection:CONNection:CATalog?
Usage:
Query only
[SENSe:]CORRection:CKIT:STANdard:DATA?
<ConnectorType>, <CalKitName>,
<KitLabel>, <StandardType>, <DelayMode>[, <Port1>[, <Port2>]]
Returns the data of the related calibration standard (identified by <ConnectorType>,
<CalKitName>, <KitLabel> and <StandardType>) and - optionally - restricted to the
given port(s).
The delay parameter is returned according to the selected <DelayMode>; see
.
Query parameters:
<ConnectorType>
Name of the connector type. Use
]CORRection:CONNection:CATalog?
names.
<CalKitName>
String parameter containing the name of the calibration kit. See
"Cal kit naming conventions " under
<KitLabel>
String parameter containing the label of the calibration kit, usu-
ally the serial number.
SCPI command reference