Command reference
R&S
®
ZNA
1052
User Manual 1178.6462.02 ─ 20
CALCulate<Chn>:TRANsform:TIME:WINDow
<WindowType>
Selects the window type for filtering the data in the frequency domain prior to the time
domain transformation.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<WindowType>
RECT | HAMMing | HANNing | BOHMan | DCHebyshev
RECT - no profiling (rectangle)
HANN - normal profile (Hann)
HAMMing - low first sidelobe (Hamming)
BOHMan - steep falloff (Bohman)
DCHebyshev - arbitrary sidelobes (Dolph-Chebychev)
*RST:
HANN
Example:
See
CALCulate<Chn>:TRANsform:TIME:DCHebyshev
Manual operation:
See
CALCulate<Chn>:TRANsform:TIME:XAXis
<Unit>
Switches over between the x-axis scaling in time units or distance units.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Unit>
TIME | DISTance
TIME - x-axis scaled in time units.
DISTance - x-axis scaled in distance units (Distance = Time * c
0
* Velocity Factor).
Example:
*RST; :CALC:TRAN:TIME:STAT ON
Reset the instrument, activating a frequency sweep, and enable
the time domain transformation for the default trace.
CALC:TRAN:TIME:XAX DIST
Convert the x-axis scaling to distance units.
Manual operation:
See
7.3.1.24
CALCulate:TRANsform:DTFault...
The
CALCulate:TRANsform:DTFault...
commands set up and control a distance
to fault measurement (with option R&S
ZNA-K2).
CALCulate<Chn>:TRANsform:DTFault:DEFine
...............................................................1053
CALCulate<Chn>:TRANsform:DTFault:DELete
...............................................................1054
CALCulate<Chn>:TRANsform:DTFault:SELect
............................................................... 1054
CALCulate<Chn>:TRANsform:DTFault:CENTer
.............................................................. 1054
CALCulate<Chn>:TRANsform:DTFault:SPAN
................................................................. 1054
SCPI command reference