Command reference
R&S
®
ZNA
1447
User Manual 1178.6462.02 ─ 20
Example:
*RST; :SOURce:POWer:CORRection:NREadings 2
SOURce:POWer:CORRection:COLLect:AVERage:
NTOLerance 0.5
Increase the number of readings and reduce the power toler-
ance to improve the accuracy.
SOURce:POWer:CORRection:ACQuire PORT,1
Perform a source power calibration using port 1 as a source
port.
SOURce:POWer:CORRection:ACQuire:VERification:
RESult?
Query the result of the verification sweep. Possible response:
Ch1 1,9.9E+002
Usage:
Query only
Manual operation:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect:AVERage[:COUNt]
<NoReadings>
Sets a limit for the number of calibration sweeps in the source power calibration. The
command is valid for all channels, ports and external generators.
Equivalent command:
SOURce<Ch>:POWer<PhyPt>:CORRection:NREadings
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the limit
applies to all channels.
<PhyPt>
Calibrated port number. This parameter is ignored; the limit
applies to all sources.
Parameters:
<NoReadings>
Number of readings
Range:
1 to 100
*RST:
2
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
Manual operation:
See
"Flatness Cal – Max Iterations"
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect:AVERage:NTOLerance
<Tolerance>
Specifies the maximum deviation of the measured power from the target power of the
calibration. The command is valid for all channels and calibrated ports.
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the toler-
ance value is valid for all sources.
SCPI command reference