GUI reference
R&S
®
ZNA
322
User Manual 1178.6462.02 ─ 20
Port
Allows you to select the source for the stimulus signal.
●
"None", "Port <Pt>", or "Conv <Pt>", if
●
"Gen <i>" if "Use Generator as Source" is checked
Remote command:
CALCulate<Ch>:PARameter:MEASure
"<Trace_Name>", "<Ratio>"
CALCulate<Ch>:PARameter:SDEFine
"<Trace_Name>", "<Ratio>"
Detector
Selects the algorithm that is used to calculate the results points from the raw measure-
ment data.
For details refer to
Chapter 4.3.5.3, "Detector settings"
Remote command:
CALCulate<Ch>:PARameter:MEASure
CALCulate<Ch>:PARameter:SDEFine
[SENSe<Ch>:]SWEep:DETector:TIME
5.2.5
Noise Figure tab
Noise figure measurements are available for both non-frequency converting and fre-
quency converting DUTs.
The noise figure measurement is provided with software option R&S
ZNA-K30. If this
option is not installed, the "Noise Figure" tab is not available.
For background information, see
Chapter 4.7.11, "Noise figure measurement"
Access:
Trace – [Meas] > "Noise Figure"
5.2.5.1
Controls on the Noise Figure tab
The controls on the "Noise Figure" tab allow you to set up a noise figure (NF) measure-
ment and give access to the related measured quantities.
Meas softtool