Command reference
R&S
®
ZNA
1314
User Manual 1178.6462.02 ─ 20
Example:
See
[SENSe<Ch>:]CORRection:EDELay<PhyPt>:ELENgth
Manual operation:
See
"Loss at DC / Loss at 1st Freq / 1st Freq for Loss / Use 2nd
Freq / Loss at 2nd Freq / 2nd Freq for Loss"
[SENSe<Ch>:]CORRection:LOSS<PhyPt>:SECond:STATe
<Boolean>
Defines whether the second reference frequency
f
2
approxi-
mation.
The second reference frequency and its associated loss value can be defined using
[SENSe<Ch>:]CORRection:LOSS<PhyPt>:SECond:FREQuency
and
[SENSe<Ch>:]CORRection:LOSS<PhyPt>:SECond
.
Suffix:
<Ch>
.
Channel number
<PhyPt>
Physical port number
Parameters:
<Boolean>
OFF (0)
f
2
not used (default)
ON (1)
f
2
used
Manual operation:
See
"Loss at DC / Loss at 1st Freq / 1st Freq for Loss / Use 2nd
Freq / Loss at 2nd Freq / 2nd Freq for Loss"
[SENSe<Ch>:]CORRection:NFIGure[:STATe]
<Boolean>
Disables or enables the Noise Figure Calibration for the related channel.
Suffix:
<Ch>
.
Parameters:
<Boolean>
OFF - Disable calibration
ON - (re-)enable calibration
*RST:
OFF (calibration disabled, no noise figure calibra-
tion available after a reset)
Options:
R&S
ZNA-K30
[SENSe<Ch>:]CORRection:OFFSet<PhyPt>[:STATe]
<Boolean>
Resets the offset parameters for all test ports to zero and the reference frequency to 1
GHz or queries whether any of the offset parameters are different from zero.
Suffix:
<Ch>
.
Channel number of the offset-corrected channel
SCPI command reference