Command reference
R&S
®
ZNA
1408
User Manual 1178.6462.02 ─ 20
[SENSe:]UDSParams:ACTive
<arg0>
With a "set to OFF", this command disables user-defined physical ports (see
ter 4.3.1.2, "Redefined S-parameters"
on page 124). In "get direction" it queries
whether or not user-defined phasical ports are configured.
Note
that a "set to ON" doesn't have any effect; user defined ports can ony be created
Parameters:
<arg0>
OFF (0) or ON (1)
Example:
Manual operation:
See
[SENSe:]UDSParams<Pt>:PARam
<PortString>
Creates or redefines a user-defined physical port (see
Chapter 4.3.1.2, "Redefined S-
on page 124). This causes a factory reset.
Suffix:
<Pt>
.
Number of the redefined physical port.
Parameters:
<PortString>
String representation '<Source>:<a wave>:<b wave>' where
<Source> =
i
refers to (original) VNA ports
<Source> = G
i
refers to external generator ports
<a wave>, <b wave> is one of the original a
i
or b
i
,
without reusing the same port in different (redefined) ports
Example:
SENSe:UDSParams:ACTive OFF
resolves all redefined ports
SENSe:UDSParams1:PARam '1:b2:b3'
redefines physical port 1 as source = (original) physical port 1,
reference receiver = physical port 2, measurement receiver =
physical port 3
SENSe:UDSParams:ACTive?
)
Manual operation:
See
7.3.15
SOURce commands
The
SOURce...
commands affect the source settings of the R&S
ZNA.
7.3.15.1
SOURce:CMODe...
The
SOURce:CMODe
(R&S
ZNA-K6).
SCPI command reference