Command reference
R&S
®
ZNA
1124
User Manual 1178.6462.02 ─ 20
CONTrol<Ch>:GPIO<Port>:VOLTage[:DEFault]
<Voltage>
Sets the output voltage of the respective GPIO port.
This voltage is applied using
CONTrol<Ch>:GPIO:VOLTage:OUTPut
Suffix:
<Ch>
.
Channel number
<Port>
GPIO port number
Parameters:
<Voltage>
Range:
-7 to +15 V
Increment: 5 mV
Default unit: V
Manual operation:
See
CONTrol<Ch>:RFFE:TEST:OUTPut
Applies the current output voltage and current range (shunt) settings to the RFFE/
GPIO pins.
CONTrol<Ch>:GPIO:VOLTage:OUTPut
.
Suffix:
<Ch>
.
Channel number
Usage:
Event
Options:
R&S ZN-B15/Z15 Var. 03
Manual operation:
See
CONTrol<Ch>:RFFE:TEST:TIME
<MeasTime>
Sets the measurement time for the voltage/current measurements on the RFFE and
GPIO pins.
Same functionality as
Suffix:
<Ch>
.
Channel number
This suffix is ignored and can be omitted: the measurement time
is valid for all channels.
Parameters:
<MeasTime>
Measurement (= sampling = averaging) time
Range:
95 μs to 100 ms
*RST:
100 ms
Default unit: s
Options:
R&S ZN-B15/Z15 Var. 03
Manual operation:
See
SCPI command reference