Command reference
R&S
®
ZNA
956
User Manual 1178.6462.02 ─ 20
The coupon type can be selected using
.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Parameters:
<String>
Path to the Touchstone file, either absolute or relative to the
selected working directory (
Manual operation:
See
"Load File / 1x Open Preset / 1x Short Preset"
CALCulate:FMODel:SFD<Ph_pt>:COUPon:MEASure:FILename:CLEar
Invalidates the file path previously set using
.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Usage:
Event
Manual operation:
See
"Load File / 1x Open Preset / 1x Short Preset"
CALCulate:FMODel:SFD<Ph_pt>:COUPon[:STATe]
<Boolean>
For a fixture modeling with the SFD tool, this command allows to specify the ports to
which the test coupon is connected.
This has to be defined before measuring the test coupon (using
).
Suffix:
<Ph_pt>
.
Physical port number
Parameters:
<Boolean>
1 (ON, true) if the test coupon is connected to port <Ph_pt>, 0
(OFF, false) otherwise
Manual operation:
See
CALCulate:FMODel:SFD<Ph_pt>:COUPon:TYPE
<CouponType>
Sets/gets the coupon type to be measured for a fixture modelling with the SFD tool.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Parameters:
<CouponType>
SYMMetric2x | OPEN1x | SHORt1x
Manual operation:
See
SCPI command reference