Command reference
R&S
®
ZNA
1393
User Manual 1178.6462.02 ─ 20
<Seg>
Sweep segment number
Parameters:
<SegPoint>
Number of points in the segment
Range:
1 to 100001. 1 is allowed if start and stop frequen-
cies are equal.
*RST:
51
Example:
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using
default settings.
SEGM:SWE:POIN 401
Increase the number of points to 401.
Manual operation:
See
[SENSe<Ch>:]SEGMent<Seg>:SWEep:TIME
<SegSweepTime>
Sets the duration of the sweep in sweep segment no.
<Seg>
("Segment Time"). At the
same time, the command activates separate sweep time setting in all sweep segments
(
[SENSe<Ch>:]SEGMent<Seg>:SWEep:TIME:CONTrol
ON
).
Suffix:
<Ch>
.
Channel number
<Seg>
Sweep segment number
Parameters:
<SegSweepTime>
Segment time. The minimum segment time depends on the
other channel settings, in particular on the number of points
(
[SENSe<Ch>:]SEGMent<Seg>:SWEep:POINts
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:
) and the delay for each partial measurement
(
[SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl
mum is 1000 s. Changing the duration leaves the number of
points unchanged but directly affects the delay.
Range:
Minimum value to 1000 s
*RST:
Minimum value, depending on the channel settings.
This default value corresponds to automatic sweep
time setting in manual control.
Default unit: s
Example:
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1.
SEGM:SWE:TIME 0.1
Increase the segment sweep time to 0.1 s.
SEGM:SWE:TIME:SUM?
Query the total duration of the segmented sweep. The response
is 0.1, because there is only one sweep segment.
Manual operation:
See
SCPI command reference