Command reference
R&S
®
ZNA
1225
User Manual 1178.6462.02 ─ 20
PROGram[:SELected]:NAME
<Program>
Selects the application to be run on the analyzer. At present, only the general parame-
ter PROG is available. This means that
any program.
Tip:
Use this command in order to avoid problems should the default value change in
future firmware versions.
Parameters:
<Program>
PROG
Any program running under Windows or any file that can be
opened with an application program available on the analyzer.
*RST:
PROG
Example:
See
PROGram[:SELected]:RETVal?
Queries the return value of an application or process started via
.
This will only be successful if preceded by a
PROGram[:SELected]:WAIT?
query
Example:
See
Usage:
Query only
PROGram[:SELected]:WAIT
Locks command execution from the current controller program while a program started
via
is running. The analyzer does not execute any
further commands or queries until the program is stopped or paused.
Use
PROGram[:SELected]:WAIT?
before trying to retrieve the return value of the
).
Example:
See
7.3.14
[SENSe:] commands
The
[SENSe:]...
commands affect the receiver settings of the R&S
ZNA.
7.3.14.1
[SENSe:]AVERage...
The
[SENSe:]AVERage...
commands set sweep averaging parameters. The sweep
average is a noise-reduction technique which consists of calculating each measure-
ment point as an average of the same measurement point over several consecutive
sweeps.
SCPI command reference