GUI reference
R&S
®
ZNA
769
User Manual 1178.6462.02 ─ 20
The table displays the faults of the active trace (in case it is a Distance to Fault
trace) .
Export to File... ← Contents
Opens a dialog that allows to export the (filtered) content of the fault list to an ASCII file
with configurable "Field Separator" and "Decimal Separator".
This CSV type file contains one line (terminated by CR+LF) per fault. With ";" as "Field
Separator" and "." as "Decimal Separator", the fault list of
is exported to
1; Trc1; 457.2 mm; -8.56 dB
2; Trc1; 1.539 m; -9.24 dB
For an explanation of the possible filters, see
Tracking Enabled
on page 768 is cleared at every sweep start and
populated in real time. Uncheck "Tracking Enabled" to freeze the table contents.
5.18
Display softtool
The "Display" softtool provides all display settings and the functions for activating,
modifying and arranging different diagrams.
Access:
System – [Display]
Related information
Refer to the following sections:
●
Chapter 4.1.3, "Traces, channels and diagrams"
●
Chapter 4.2.1, "Display elements of a diagram"
●
Chapter 3.3.5, "Handling diagrams, traces, and markers"
Display softtool