Command reference
R&S
®
ZNA
1171
User Manual 1178.6462.02 ─ 20
This command is not supported in the "ZVR" or "ZVABT" compatibility modes.
Note:
In contrast to all other commands of the analyzer, the
INITiate<Ch>[:IMMediate]...
commands have been implemented for overlap-
Chapter 6.4, "Command processing"
Example:
See
Usage:
Event
Manual operation:
See
INITiate:CONTinuous:ALL
<Boolean>
Qualifies whether the analyzer measures in single sweep or in continuous sweep
mode.
This command is not supported in the "ZVR" or "ZVABT" compatibility modes.
Parameters:
<Boolean>
ON - the analyzer measures continuously in all channel, repeat-
ing the current sweep. The query returns ON (1) if at least one
channel is measured continuously.
OFF - the measurement is stopped after the number of sweeps
defined via
initiates a new measurement cycle.
*RST:
ON
Example:
See
Manual operation:
See
"All Channels Continuous/All Channels on Hold"
INITiate<Ch>[:IMMediate][:DUMMy]
Starts a new single sweep sequence. This command is available in single sweep mode
only (
OFF
). The data of the last sweep (or previous
sweeps, see
Chapter 8.2.4.3, "Retrieving the results of previous sweeps"
on page 1613) can be read using
CALCulate<Chn>:DATA:NSWeep:FIRSt?
SDATa, <count>
.
Note:
In contrast to all other commands of the analyzer, the
INITiate<Ch>[:IMMediate]...
commands have been implemented for overlap-
Chapter 6.4, "Command processing"
Suffix:
<Ch>
.
Channel number. If the channel number does not exist the ana-
lyzer returns an error message. If the "ZVR" or "ZVABT" compat-
ibility mode is active and
ALL
is selected, this suffix is ignored.
Example:
See
Usage:
Event