Command reference
R&S
®
ZNA
1292
User Manual 1178.6462.02 ─ 20
Manual operation:
See
"Same Connector All Ports / Same Gender All Ports"
[SENSe<Ch>:]CORRection:COLLect:CONNection:PORTs
<ConnectorType>
Qualifies whether the connector types at the test ports (but not their gender) are equal
or independent. Some calibration types require uniform port connector types.
Suffix:
<Ch>
.
Channel number of the calibrated channel
Parameters:
<ConnectorType>
ALL | SINGle
ALL
–
equal (uniform) connector types. If the connector type at
one port is changed, the connector type at all other ports is
changed accordingly.
SINGle
–
independent (possibly non-uniform) connector types at
the ports.
*RST:
ALL
Example:
CORR:COLL:CONN:PORTS SING
Select independent connector types at the ports.
CORR:COLL:CONN1 N50MALE; CONN4 N75FEMALE;
CONN2?
Select independent connector types at ports 1 and 4. The con-
nector type at port 2 is not changed; the query returns
N50FEMALE
.
Manual operation:
See
"Same Connector All Ports / Same Gender All Ports"
[SENSe<Ch>:]CORRection:COLLect:CSETup
<Boolean>
Defines how calibration sweeps are prepared and performed.
This setting is valid for manual and automatic calibration.
Suffix:
<Ch>
.
Channel number (ignored)
Parameters:
<Boolean>
OFF (0)
Individual sweep setups for all standards (faster for some config-
urations)
ON (1)
Common sweep setup for all standards (no preparation phase
for each standard)
*RST:
OFF
Manual operation:
See
"Same Sweep Setup for All Standards"
SCPI command reference