Command reference
R&S
®
ZNA
1400
User Manual 1178.6462.02 ─ 20
Parameters:
<Value>
*RST:
1 ms
Default unit: s
Options:
R&S
ZNA-K28
Manual operation:
See
[SENSe<Ch>:]SWEep:CTIMing:STIMe:MODE
<Boolean>
Enables/disables user-defined wait times per measurement point for the controlled tim-
ing mode (
.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean>
OFF (0)
Wait time is calculated by the analyzer firmware
ON (1)
Wait time can be specified manually using
Options:
R&S
ZNA-K28
Manual operation:
See
[SENSe<Ch>:]SWEep:DETector:TIME
<DetectorTime>
Defines the observation time per sweep point if a detector other than the "Normal"
(SAMple) detector is used. The detector is selected together with the measured quan-
tity (
CALCulate<Ch>:PARameter:MEASure
).
Suffix:
<Ch>
.
Channel number
Parameters:
<DetectorTime>
Detector time
Range:
0 s to 3456000 s
*RST:
0.01 s
Default unit: s
Example:
*RST; :CALC:PAR:MEAS 'TRC1', 'a1D1AVG'
Select the wave quantity a
1
for the default trace and activate the
AVG detector.
SWE:DET:TIME 1
Specify an observation time of 1 s at each sweep point.
Manual operation:
See
SCPI command reference