GUI reference
R&S
®
ZNA
563
User Manual 1178.6462.02 ─ 20
If option K50 "Measurement Uncertainty Analysis" is installed, the "KitModified" column
is replaced by an "Uncertainty" column, indicating whether the kits contains uncertainty
information for all its calibration standards.
Remote command:
[SENSe:]CORRection:CKIT:CATalog?
Add / Copy / Delete / Standards...
The buttons in the right part of the dialog are used to manage calibration kits:
●
"Add" creates a cal kit file for the selected connector type.
●
"Copy" creates a cal kit file based on the contents of an existing cal kit file.
●
"Delete" deletes an imported or user-defined cal kit file.
●
"Standards..." opens the "Kit Standards" dialog, which shows the contents of the
cal kit file. For user-defined or imported kits, you can modify the contents. See
Remote command:
The following commands create and modify calibration kits:
[SENSe:]CORRection:CKIT:<ConnType>:SELect
[SENSe:]CORRection:CKIT:<ConnType>:LSELect
[SENSe:]CORRection:CKIT:DMODe
[SENSe:]CORRection:CKIT:DELete
Query connector types and calibration kits:
[SENSe<Ch>:]CORRection:CONNection:CATalog?
[SENSe:]CORRection:CKIT:CATalog?
Import Cal Kit... / Export Cal Kit...
The buttons below the "Connector Type" list are used to store cal kit data to a file and
to reload previously stored cal kit files. By default, calibration kit files are stored in the
C:\Users\Public\Documents\Rohde-Schwarz\ZNA\Calibration
directory;
Chapter 4.5.2.4, "Cal kit Files"
Remote command:
MMEMory:LOAD:CKIT
MMEMory:STORe:CKIT
Kit Standards dialog
The "Kit Standards" dialog shows the calibration standards in a selected calibration kit.
It is also used to modify the contents of a user-defined kit.
Related information
Refer to the following sections:
●
Chapter 4.5.2.4, "Cal kit Files"
●
Chapter 5.11.2.2, "Calibration Kits dialog"
●
Chapter 4.5.2.1, "Calibration standard types"
Access:
Cal softtool