R&S
®
ZNA
1437
User Manual 1178.6462.02 ─ 20
SOURce:POWer<PhyPt>:CORRection:SLEVeling:DELete
Removes the (global) leveling dataseet of port <PhyPt>.
Suffix:
<PhyPt>
.
Physical port number
Usage:
Event
Options:
R&S
ZNA-K8
Manual operation:
See
SOURce:POWer<PhyPt>:CORRection:SLEVeling:STATe?
Queries whether a
data set is available for physical port <PhyPt>.
Suffix:
<Ch>
.
Channel number
This suffix is ignored. Leveling data are valid for all channels in
the active setup.
Return values:
<Boolean>
Usage:
Query only
Options:
R&S
ZNA-K8
Manual operation:
See
SOURce:POWer<PhyPt>:CORRection:SLEVeling:FREQuency:POINts
<NrOfPts>
Defines the number of frequency sweep points of the
grid for physical port
<PhyPt>.
After defining the leveling grid (using
SOURce<Ch>:POWer<PhyPt>:CORRection:SLEVeling:FREQuency
and
SOURce<Ch>:POWer<PhyPt>:CORRection:SLEVeling:POWer
commands), you
can start the leveling data acquisition using
SOURce:POWer<PhyPt>:CORRection:
.
Suffix:
<PhyPt>
.
Physical port number
Parameters:
<NrOfPts>
Number of points
By default the R&S
ZNA uses 201 sweep points. The higher the
number of "Sweep Points", the more accurate the leveling, but
also the longer it takes to create the leveling table). The number
of sweep points is limited to 100001.
Example:
See
SOURce:POWer<PhyPt>:CORRection:SLEVeling[:
Options:
R&S
ZNA-K8