Command reference
R&S
®
ZNA
1282
User Manual 1178.6462.02 ─ 20
Setting parameters:
<CalName>
Example:
SENSe1:CORRection:COLLect:AUTO:CONFigure
FNPort,''
Sets the automatic calibration to "Full n-port" with factory char-
acterization.
Example:
See
Chapter 8.2.5.6, "Vector mixer calibration using calibration
Manual operation:
See
[SENSe<Ch>:]CORRection:COLLect:AUTO:CPORt
<PortNumber>
Defines the "Common Port" of an automatic calibration.
This port will be part of every port assignment that is defined using
]CORRection:COLLect:AUTO:ASSignment:DEFine:DEFault
]CORRection:COLLect:AUTO:ASSignment:DEFine:TPORt:DEFault
If
[SENSe:]CORRection:COLLect:AUTO:MCONnect
ON
, this port will be used as
the center of all Through measurements. If you set up the port assignments manually,
make sure that every port assignment contains it.
Suffix:
<Ch>
.
Channel number
Setting parameters:
<PortNumber>
Number of the common port
Usage:
Setting only
Manual operation:
See
[SENSe<Ch>:]CORRection:COLLect:AUTO:MIXer
<Port1>, <Port2>
Acquire calibration data for the calibration mixer in an automatic vector mixer calibra-
tion
Suffix:
<Ch>
.
Channel number
Setting parameters:
<Port1>
VNA port connected to the RF port of the calibration mixer
<Port2>
VNA port connected to the IF port of the calibration mixer
Example:
See
Chapter 8.2.5.6, "Vector mixer calibration using calibration
Usage:
Setting only
Options:
R&S
ZNA-K5
SCPI command reference