Command reference
R&S
®
ZNA
1108
User Manual 1178.6462.02 ─ 20
Query parameters:
<Data>
ALL
If omitted, a single numeric value is returned. If ALL is specified,
the result consists of 6 numeric values. Furthermore, the inter-
pretation of the result values depends on the current trace's
stimulus axis (
CALCulate<Chn>:TRANsform:TIME:XAXis
See the table below.
Usage:
Query only
Options:
R&S
ZNA-K20
Manual operation:
See
<Data> omitted
<Data> = ALL
Stimulus axis:
time
<rise time>
<rise time>,<rise distance>, <start threshold crossing time>,
<start threshold crossing voltage>, <stop threshold crossing
time>, <stop threshold crossing voltage>
Stimulus axis:
distance
<rise distance>
<rise distance>, <rise time>,<start threshold crossing dis-
tance>, <start threshold crossing voltage>, <stop threshold
crossing distance>, <stop threshold crossing voltage>
CALCulate<Chn>:TTIMe:STATe
<Boolean>
Enables/disables the
Note:
The rise time measurement can only be enabled if the active trace is real
REAL
), Time Domain is enabled (
ON
), and the Low Pass Step time domain transform is used
CALCulate<Chn>:TRANsform:TIME[:TYPE]
LPASs
and
STEP
). The latter, in turn, requires the stimulus grid to
be harmonic. This can be achieved, for example, using
ON
.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean>
ON – Rise time measurement enabled
OFF – Rise time measurement disabled
*RST:
OFF
Options:
R&S
ZNA-K20
Manual operation:
See
CALCulate<Chn>:TTIMe:THReshold
<ThresholdEnum|LowerThreshold>[,
<UpperThreshold>]
Defines the lower/upper threshold for the rise time measurement.
SCPI command reference