Important information before getting started

 

First use: Before the initial operation of JuLI™ Stage, we recommend to 

leave the instrument outside an incubator for 20 minutes, keeping the power 
on. This will allow the instrument’s temperature to equilibrate to room 

temperature. 

 

If the instrument is to be operated inside an incubator, additionally warm up 
the instrument to 37 °C for 2 hours. Make sure to keep the power on during 
this process. 

 

Environmental requirements: When operating JuLI™ Stage inside an 
incubator, make sure to keep the laboratory’s temperature in a range of 
between 15 °C and 25 °C. Temperature beyond this range may affect the 
performance of the JuLI™ Stage. 

 

Installation: Please make sure to carefully follow the installation instructions 

in this manual in order to ensure safe operation. If all the components are 
connected correctly, it will look like the Image below. 

 

PC 

USB-Repeater 

USB-Cables 

Control Box 

Connection Cable 

JuLI™ Stage 

 

 

Make sure no cable is under tension. 

 