Keys

Functions

Powers the meter on and off . Upon power on, the meter 

automatically begins in the mode that was last used. Calibration 

and memory values are retained even if meter is unplugged.
Toggle between available measurement modes; pH/Temp, 

mV/Temp, or ppm/mV (Ion 700 only). 

Also used to switch from pH to Temp during pH calibration 

mode. 

Press and hold for 5 seconds to enter SETUP mode.

Toggles between measurement and calibration modes. 

In SETUP mode, returns user to the measurement mode.

MI

 (Memory Insert) stores values into memory.

 Increase value. Scroll up in SETUP mode.

MR

 (Memory Recall) recalls values from memory

 Decrease value. Scroll down in SETUP mode.

Freezes measured reading. Press again to resume live reading.

Confi rms calibration values in CAL mode. 

Confi rms selections in SETUP mode.

View recalled values in memory mode.

n

  

LCD Annunciators

Quick Guide 

Bench 700 Series

pH/mV/Ion/Conductivity/TDS/Dissolved Oxygen/ºC/ºF

n

  

Keypad Functions

Part of Thermo Fisher Scientifi c

The 700 series meters can retain up to 100 points into memory for later retrieval. 

1.  In the measurement mode, press 

 to insert the measured value into memory. 

The stored memory location value (StO) is briefl y displayed.

2.  To recall data from memory, press 

. The location of the most recent stored data is displayed 

fi rst. Press 

 or 

 to select the location of the desired data, then press 

 to accept.

3. Press 

 to return to the stored data location. Press 

 to return to measurement mode. 

For more information on our products, please contact our channel partner or visit our websites listed below: 

Eutech Instruments Pte Ltd  Oakton Instruments 

Eutech Instruments Europe B.V.

Blk 55, Ayer Rajah Crescent, 

625 E Bunker Ct,  

P.O. Box 254, 3860 AG Nijkerk,

#04-16/24, Singapore 139949

Vernon Hills, IL 60061, USA

The Netherlands Wallerstraat 125K,

eutech@thermofi sher.com  

[email protected] 

3862 CN Nijkerk, The Netherlands

www.eutechinst.com  

www.4oakton.com 

eutech@thermofi sher.com

 www.eutechinst.com

Automatic 

temperature 

compensation 

is active

Setup 

mode

Measured 

value is held

Current 

measurement 

unit / mode 

of operation

Electrode error 

when appears 

with ERR

Primary 

reading

Secondary

reading

During pH CAL, 

secondary reading is 

the auto recognized pH 

buff er standard value

Measured 

value is stable

Error or invalid 

key pressed

Prompts user 

to select buff er

During setup, 

select auto hold 

ON, ready ON or 

Ready OFF

Measurement

mode

Calibration

mode

Memory mode or 

when data is stored 

into memory

GETTING STARTED

STORING AND RECALLING DATA

n

Ion Calibration (Ion 700 Only)

The available ION calibration values are 0.10, 1.0, 10.0, 100.0, and 1000 parts per million (ppm). 

Choose a minimum of 2 consecutive values for calibration and prepare the corresponding ion 

calibration solutions. For best results always begin with your lowest standard value, followed by 

the next lowest standard. 

1.  Connect the ISE and press 

 as needed to select ppm. Note: The primary reading will show  

“- - - -” without a stored ion calibration. The secondary value is the corresponding mV reading 
of the ISE. 

2.  Dip the ISE into your lowest standard solution value and stir. Press 

 when the secondary  

reading is stable. 

3. Press 

 or 

 to match the primary display to your corresponding ion calibration 

value (0.10, 1.0, 10, etc). 

4. Press 

 to accept or 

 to abort. The primary display will show the next highest calibration   

standard value.

5.  Rinse the ISE then dip into the corresponding calibration standard and stir.  

6. Press 

 to accept or 

 to cancel. The mV/decade slope value will be displayed briefl y if the  

calibration is successful. “SLP Err” indicates calibration for the current point was not successful. 

7. Revert to step 3 to calibrate additional points or press 

 as needed to return to measurement 

mode at any time. 