Command reference
R&S
®
ZNA
1352
User Manual 1178.6462.02 ─ 20
Example:
SENSe1:NFIGure:DEFine 2,1
Initiates a noise figure measurement in channel 1, with drive port
1 and receive port 2.
Usage:
Setting only
Options:
R&S
ZNA-K30
Manual operation:
See
"Driving Port (RF) / Receiving Port (IF)"
[SENSe<Ch>:]NFIGure:GTIMe
<DetectorTime>
Defines the detector time of the AVG detector which is used for the gain measurement
during a NF measurement in channel <Ch>
Suffix:
<Ch>
.
Channel number
Parameters:
<DetectorTime>
Detector time for the AVG detector
Default unit: s
Example:
SENSe1:NFIGure:GTIMe 10ms
Options:
R&S
ZNA-K30
Manual operation:
See
[SENSe<Ch>:]NFIGure:HARMonic[:STATe]
<State>
Activates or deactivates the sideband correction for a noise figure measurement in
channel <Ch>.
Suffix:
<Ch>
.
Channel number
Parameters:
<State>
Enables (1) or disables (0) the sideband correction of the noise
figure calibration and measurement.
*RST:
1
Options:
R&S
ZNA-K30
Manual operation:
See
[SENSe<Ch>:]NFIGure:HARMonic[:MINFrequency]
<MinimumFrequency>
[SENSe<Ch>:]NFIGure:HARMonic[:MAXFrequency]
<FrequencyLimit>
Sets the lower|upper frequency limit to which the sideband correction is used in a noise
figure measurement in channel <Ch>.
Suffix:
<Ch>
.
Channel number
SCPI command reference