GUI reference
R&S
®
ZNA
749
User Manual 1178.6462.02 ─ 20
Each value describes the jitter for a symbol transition (even when the current and the
previous symbols are the same). The jitter value is implicitly given in the unit [s] and
denotes the delta between the time of the ideally expected transition (given by the data
rate) and the actual one. Positive values describe a transition that is occurring after the
ideally expected transition. The values are limited to the range -1.0e-3 to +1.0e-3. Any
values outside this range are limited to this range.
After the values in the file have been exhausted, the jitter values are taken starting
from the beginning of the file.
Remote command:
CALCulate<Chn>:EYE:JITTer:TYPE:USER
MMEMory:LOAD:EYE:JITTer
Noise
The "Define Noise" dialog allows you to add Gaussian noise to the digital signal simu-
lated for the eye diagram.
Access:
"Noise" button in the
This dialog is available with
ZNA-K20 only.
Active
Enables/disables noise insertion.
Remote command:
CALCulate<Chn>:EYE:NOISe:STATe
RMS
Defines the root mean square noise level.
Remote command:
DUT
The "Define DUT" dialog allows you to switch between different transmission (and
reflection) paths of the DUT and to configure its DC properties. In addition, it allows
you to switch between the real DUT (with measured frequency response) and an ideal
one (with flat frequency response), which can be useful when configuring the other eye
diagram blocks (see
Chapter 5.17.2.6, "Advanced Settings dialog"
Access:
Applic softtool