Command reference
R&S
®
ZNA
961
User Manual 1178.6462.02 ─ 20
Manual operation:
See
Assume that the result at a sweep point is given by the complex quantity z = x + jy.
Then the magnitude of z is calculated as
|z| = sqrt(x
2
+ y
2
)
and in phase notation we have
z = |z| e
j Phase(z)
, where Phase(z) = arctan(y/x).
The meaning of the parameters is as follows (see also table in
description):
MLINear
Displays |z| in a Cartesian diagram
MLOGarithmic
MAGNitude (for compatibility with R&S ZVR ana-
lyzers)
Calculates |z| in dB (= 20 log|z|) and displays it in a
Cartesian diagram
PHASe
Calculates Phase(z) in the range between -180° and
+180° and displays it in a Cartesian diagram
UPHase
Calculates Phase(z) (unwrapped) and displays it in
a Cartesian diagram
POLar
COMPlex (for compatibility with R&S ZVR ana-
lyzers)
Displays z in a polar diagram
SMITh
Displays z in a Smith diagram
ISMith
Displays z in an inverted Smith diagram
GDELay
For frequency sweeps only
Calculates the group delay at the related sweep
point and displays it in a Cartesian diagram
REAL
Calculates Re(z) = x and displays it in a Cartesian
diagram
IMAGinary
Calculates Im(z) = y and displays it in a Cartesian
diagram
SWR
Calculates the standing wave ratio (1 + |z|) / (1 - |z|)
and displays it in a Cartesian diagram
LOGarithmic
Displays z in a Cartesian diagram with logarithmic
scale
CALCulate<Chn>:FORMat:WQUType
<Unit>
Selects the physical unit of the displayed trace.
Suffix:
<Chn>
.
Channel number used to identify the active trace
SCPI command reference