SENSe1:SEGMent:POWer:GAINcontrol:CONTrol OFF
// config similar to the one done inside NF Meas Dialog
// define ports for NF
SENSe1:NFIGure:DEFine 2, 1 // receiving port 2, driving port 1
// set stimulus
SENSe1:FREQuency:STARt 100MHz
SENSe1:FREQuency:STOP 4GHz
SENSe1:SWEEp:POInts 40
// base power (for DUT Measurement)
SOURce1:POWer -30dBm
// set RBW (2MHz equivalent bandwidth)
SENS1:BAND:RES 2MHz
// set source step attenuator to 40dB at source port
SOURce1:POWer1:ATT 40dB
// use LNA with 25dB gain at port 2
SENSe1:PAMPlifier2:STATe ON
SENSe1:PAMPlifier2:VALue 25
// noise detector measurement time (for measurement)
SENSe1:NFIGure:NTIMe 50ms
// gain detector measurement time (for measurement)
SENSe1:NFIGure:GTIMe 10ms
// noise detector measurement time (for calibration)
SENSe1:NFIGure:CALibration:NTIMe 50ms
// gain detector measurement time (for calibration)
SENSe1:NFIGure:CALibration:GTIMe 10ms
// configure harmonic correction
:SENSe1:NFIGure:HARMonic:STATe 1
:SENSe1:NFIGure:HARMonic:MAX 5
:SENSe1:NFIGure:HARMonic:MINFrequency 10MHz
:SENSe1:NFIGure:HARMonic:MAXFrequency 20GHz
// reference receiver power calibration: calibration power level of 0dBm
SOURce1:POWer:CORRection:PPOWer 0dBm
SOURce1:POWer:CORRection:PSELect PPOWer
// value for source step attenuator is 0dB by default
SOURCe1:POWer:CORRection:PPOWer:PATTenuation 0dB