Command reference
R&S
®
ZNA
1445
User Manual 1178.6462.02 ─ 20
SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
<SourceType>[,
<SourcePort>[, <CalOnlyPortFreq>]]
Selects the source for the source power calibration, starts and applies the source
power calibration.
Note:
The command cannot be used unless a power meter is connected via GPIB bus,
USB or LAN interface and configured in the "External Power Meters" dialog.
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt>
Calibrated port number. This suffix is ignored because the port
number is specified in the parameter list.
Setting parameters:
<SourceType>
PORT | GENerator | A1 | A2 | A3 | A4 | ESRC1 | ESRC2 | CLO
PORT
–
analyzer port is the source of the calibrated wave.
GENerator
–
external generator is the source.
The port and generator numbers are specified by means of the
<SourcePort>
parameter.
A1 | A2 | A3 | A4
–
direct analyzer port selection. The parame-
ters A1 and A2 are ZVR-compatible. No additional port number
parameter needs to be specified.
ESRC1 | ESRC2
–
direct external generator selection. The
parameters are ZVR-compatible. No additional parameters need
to be specified.
CLO
–
port.
<SourcePort>
Analyzer port or generator port number, if PORT or GENerator is
the first parameter.
<CalOnlyPortFreq>
Relevant for frequency conversion measurements only.
OFF | 0
The reference receiver at port <CalPort> is calibrated for all fre-
quencies that are relevant for any of the ports.
ON | 1
Only those frequencies are calibrated that are relevant for this
port. This results in shorter calibration times.
*RST:
0
SCPI command reference